Specifications
The Biohit eLINE Electronic Pipettors may be ordered with or without an AC-adaptor. When ordered without the AC-adaptor the eLINE Charging Stand or Carousel is necessary.
| Biohit eLINE Electronic Pipettors | ||||||||||
| Cat.No. | Ch. | Volume range |
Increment | Test volume |
Inacc. | Impr. | Safe-Cone Filters | Tip ID | ||
| Standard | Plus | |||||||||
| 73002X | 1-ch | 0.2-10 µl | 0.05 µl | 10 µl 5 µl 1 µl 0.2µl |
0.90 % 1.00 % 2.50 % 12.00 % |
0.40% 0.70% 1.50% 10.00% |
N/A | N/A | A L,M |
|
| 73004X | 1-ch | 5-120 µl | 0.50 µl | 120 µl 60 µl 12 µl 5 µl |
0.40 % 0.60 % 2.00 % 3.50 % |
0.15% 0.20% 1.00% 1.50% |
721008 | 721018 | C,D V | |
| 73006X | 1-ch | 10-300 µl | 1.00 µl | 300 µl 150 µl 30 µl 10 µl |
0.40 % 0.60 % 1.50 % 3.50 % |
0.15% 0.20% 0.80% 2.00% |
721007 | 721017 | D Q |
|
| 73008X | 1-ch | 50-1000 µl | 5.00 µl | 1000 µl 500 µl 100 µl 50 µl |
0.40 % 0.60 % 1.50 % 2.00 % |
0.15% 0.20% 0.50% 1.00% |
721006 | 721016 | E,G S | |
| 73010X | 1-ch |
100-5000 µl * |
10.0 µl | 5000 µl 2500 µl 500 µl |
0.50 % 0.80 % 1.00 % |
0.15% 0.20% 0.40% |
721006 | 721016 | J | |
| 73032X | 8-ch | 0.2-10 µl | 0.05 µl | 10 µl 5 µl 1 µl |
0.90 % 1.50 % 4.00 % |
0.50% 0.80% 3.00% |
N/A | N/A | A L,M |
|
| 73034X | 8-ch | 5-120 µl | 0.50 µl | 120 µl 60 µl 12 µl |
0.50 % 0.70 % 2.00 % |
0.20% 0.30% 1.50% |
721008 | 721018 | C,D V | |
| 73036X | 8-ch | 10-300 µl | 1.00 µl | 300 µl 150 µl 30 µl |
0.50 % 0.70 % 2.00 % |
0.20% 0.30% 1.00% |
721007 | 721017 | D Q | |
| 73039X | 8-ch | 50-1200 µl | 5.00 µl | 1200 µl 600 µl 120 µl |
0.50 % 1.00 % 2.50 % |
0.20% 0.30% 1.00% |
721006 | 721016 | H T | |
| 73042X | 12-ch | 0.2-10 µl | 0.05 µl | 10 µl 5 µl 1 µl |
0.90 % 1.50 % 4.00 % |
0.50% 0.80% 3.00% |
N/A | N/A | A L,M |
|
| 73044X | 12-ch | 5-120 µl | 0.50 µl | 120 µl 60 µl 12 µl |
0.50 % 0.70 % 2.00 % |
0.20% 0.30% 1.50% |
721008 | 721018 | C,D V | |
| 73046X | 12-ch | 10-300 µl | 1.00 µl | 300 µl 150 µl 30 µl |
0.50 % 0.70 % 2.00 % |
0.20% 0.30% 1.00% |
721007 | 721017 | D Q | |
| 73049X | 12-ch | 50-1200 µl | 5.00 µl | 1200 µl 600 µl 120 µl |
0.50 % 1.00 % 2.50 % |
0.20% 0.30% 1.00% |
721006 | 721016 | H T | |
|
X: 0 = Without AC-adaptor N/A = Not available | ||||||||||