Decontaminating eLINE

Daily Safety Maintenance:

Lightly spray the outside of the pipettor and wipe dry with a clean cloth.

Complete eLINE pipettor decontamination procedure:

1. eLine Contamination ProductsUnscrew the tip ejector collar counterclockwise and remove it.
2. eLine Decontamination ProductsUnscrew the tip cone holder counterclockwise and carefully remove the tip cone holder, tip cone and spring. Remove the Safe-Cone Filter if fitted.
3. eLine Decontamination ProductsUnscrew the piston counterclockwise from the pipettor.
4. eLine Decontamination ProductsPlace the tip cone, tip cone holder, tip ejector collar, piston and spring into a beaker containing Biohit Proline Biocontrol and leave for at least 30 minutes.
5. After performing the procedure described above, remove the components from the beaker and rinse them with distilled water, then dry preferably with warm air, for at least an hour.
6. Re-grease the piston as described in the instruction manual. Replace all components including new filter if fitted.

Note: When performing the decontamination procedure as a matter of routine, the o-ring seal should be checked for possible wear at every sixth instance, and replaced if necessary. The calibration should also be checked at the same time.