mLINE - the most ergonomic choice




| Item | Channels | Volume | Increment | Test volume | Inaccuracy | Imprecision |
| mLINE Mechanical Pipette | 1 | 0.1-3 µl | 0.002 µl | 3 µl 1.5 µl 0.3 µl |
1.30 % 2.40 % 10.00 % |
0.80 % 1.60 % 6.00 % |
| mLINE Mechanical Pipette | 1 | 0.5-10 µl | 0.01 µl | 10 µl 5 µl 1 µl 0.5 µl |
1.00 % 1.50 % 2.50 % 5.00 % |
0.60 % 1.00 % 1.50 % 4.00 % |
| mLINE Mechanical Pipette | 1 | 2-20 µl | 0.02 µl | 20 µl 10 µl 2 µl |
0.90 % 1.20 % 3.00 % |
0.40 % 1.00 % 2.00 % |
| mLINE Mechanical Pipette | 1 | 10-100 µl | 0.10 µl | 100 µl 50 µl 10 µl |
0.80 % 1.00 % 2.00 % |
0.15 % 0.40 % 1.00 % |
| mLINE Mechanical Pipette | 1 | 20-200 µl | 0.20 µl | 200 µl 100 µl 20 µl |
0.60 % 0.80 % 2.00 % |
0.15 % 0.30 % 0.80 % |
| mLINE Mechanical Pipette | 1 | 100-1000 µl | 1.00 µl | 1000 µl 500 µl 100 µl |
0.60 % 0.60 % 1.00 % |
0.20 % 0.20 % 0.40 % |
| mLINE Mechanical Pipette | 1 | 500-5000 µl | 10.0 µl | 5000 µl 2500 µl 500 µl |
0.50 % 0.60 % 2.00 % |
0.20 % 0.30 % 0.60 % |
| mLINE Mechanical Pipette | 1 | 1-10 ml | 20.0 µl | 10000 µl 5000 µl 1000 µl |
0.60 % 1.20 % 3.00 % |
0.20 % 0.30 % 0.60 % |
| mLINE Mechanical Pipette | 8 | 0.5-10 µl | 0.01 µl | 10 µl 5 µl 1 µl |
1.50 % 2.50 % 4.00 % |
1.00 % 2.50 % 4.00 % |
| mLINE Mechanical Pipette | 8 | 5-100 µl | 0.10 µl | 100 µl 50 µl 10 µl |
0.70 % 1.00 % 3.00 % |
0.25 % 0.70 % 1.50 % |
| mLINE Mechanical Pipette | 8 | 30-300 µl | 0.20 µl | 300 µl 150 µl 30 µl |
0.60 % 1.00 % 2.00 % |
0.25 % 0.50 % 1.00 % |
| mLINE Mechanical Pipette | 12 | 0.5-10 µl | 0.01 µl | 10 µl 5 µl 1 µl |
1.50 % 2.50 % 4.00 % |
1.00 % 2.50 % 4.00 % |
| mLINE Mechanical Pipette | 12 | 5-100 µl | 0.10 µl | 100 µl 50 µl 10 µl |
0.70 % 1.00 % 3.00 % |
0.25 % 0.70 % 1.50 % |
| mLINE Mechanical Pipette | 12 | 30-300 µl | 0.20 µl | 300 µl 150 µl 30 µl |
0.60 % 1.00 % 2.00 % |
0.25 % 0.50 % 1.00 % |
| Compatible | |||||||||
| Cat. No | Channels | Volume | Non-Filtered Tip | Filtered Tip | Safe Cone Std. | Safe Cone Plus. | Stand | Quote | |
| 725010 | 1 | 0.1-3 µl | 725600 725610 725620 |
| |||||
| 725020 | 1 | 0.5-10 µl | 725600 725610 725620 |
| |||||
| 725030 | 1 | 2-20 µl | 721014 | 725600 725610 725620 |
| ||||
| 725050 | 1 | 10-100 µl | 721008 | 721018 | 725600 725610 725620 |
| |||
| 725060 | 1 | 20-200 µl | 721007 | 721017 | 725600 725610 725620 |
| |||
| 725070 | 1 | 100-1000 µl | 721006 | 721016 | 725600 725610 725620 |
| |||
| 725080 | 1 | 500-5000 µl | 721005 | 721015 | 725600 725610 725620 |
| |||
| 725090 | 1 | 1-10 ml | 721005 | 721015 | 725600 725610 725620 |
| |||
| 725120 | 8 | 0.5-10 µl | 725600 725610 725620 |
| |||||
| 725130 | 8 | 5-100 µl | 721008 | 721018 | 725600 725610 725620 |
| |||
| 725140 | 8 | 30-300 µl | 721007 | 721017 | 725600 725610 725620 |
| |||
| 725220 | 12 | 0.5-10 µl | 725600 725610 725620 |
| |||||
| 725230 | 12 | 5-100 µl | 721008 | 721018 | 725600 725610 725620 |
| |||
| 725240 | 12 | 30-300 µl | 721007 | 721017 | 725600 725610 725620 |
| |||
Accredited calibration and future demands for pipettors
Type: article
mLINE 10 ml card
Type: brochure
mLINE 3 µl card
Type: brochure
mLINE manual in English
Type: manual
mLINE manual in Finnish
Type: manual
mLINE manual in French
Type: manual
mLINE manual in German
Type: manual
mLINE manual in Italian
Type: manual
mLINE manual in Russian
Type: manual
mLINE manual in Spanish
Type: manual
mLINE manual in Swedish
Type: manual
mLINE PCR Starter kit brochure
Type: brochure
mLINE product brochure
Type: brochure
Raising The Standard of Mechanical Pipetting
Type: article
The effect of filters on aseptic pipetting lifetime of mechanical and electronic pipettors and carry
Type: article