Twice as fast as a mechanical pipette
Operation modes
Pipetting
Reverse Pipetting
Super Pipetting *
Pipetting+Mixing
Multiple Dispensing
Diluting
Diluting+Mixing
Sequential Dispening **
Multi Aspirating
* Only in eLINE 0.1-5 µl
** Excl. eLINE 0.1-5 µl




| Item | Channels | Volume | Increment | Test volume | Inaccuracy | Imprecision |
| eLINE Electronic Pipette | 1 | 0.2-10 µl | 0.05 µl | 10 µl 5 µl 1 µl 0.2 µl |
0.90 % 1.00 % 2.50 % 12.00 % |
0.40 % 0.70 % 1.50 % 10.00 % |
| eLINE Electronic Pipette | 1 | 5-120 µl | 0.50 µl | 120 µl 60 µl 12 µl 5 µl |
0.40 % 0.60 % 2.00 % 3.50 % |
0.15 % 0.20 % 1.00 % 1.50 % |
| eLINE Electronic Pipette | 1 | 10-300 µl | 1.00 µl | 300 µl 150 µl 30 µl 10 µl |
0.40 % 0.60 % 1.50 % 3.50 % |
0.15 % 0.20 % 0.80 % 2.00 % |
| eLINE Electronic Pipette | 1 | 50-1000 µl | 5.00 µl | 1000 µl 500 µl 100 µl 50 µl |
0.40 % 0.60 % 1.50 % 2.00 % |
0.15 % 0.20 % 0.50 % 1.00 % |
| eLINE Electronic Pipette | 1 | 100-5000 µl | 10.0 µl | 5000 µl 2500 µl 500 µl |
0.50 % 0.80 % 1.00 % |
0.15 % 0.20 % 0.40 % |
| eLINE Electronic Pipette | 1 | 0.1-5 µl | 0.05 | 5.0 µl 2.5 µl 0.5 µl |
1.20 % 1.70 % 9.50 % |
0.70 % 1.20 % 8.00 % |
| eLINE Electronic Pipette | 8 | 0.2-10 µl | 0.05 µl | 10 µl 5 µl 1 µl |
0.90 % 1.50 % 4.00 % |
0.50 % 0.80 % 3.00 % |
| eLINE Electronic Pipette | 8 | 5-120 µl | 0.50 µl | 120 µl 60 µl 12 µl |
0.50 % 0.70 % 2.00 % |
0.20 % 0.30 % 1.50 % |
| eLINE Electronic Pipette | 8 | 10-300 µl | 1.00 µl | 300 µl 150 µl 30 µl |
0.50 % 0.70 % 2.00 % |
0.20 % 0.30 % 1.00 % |
| eLINE Electronic Pipette | 8 | 50-1200 µl | 5.00 µl | 1200 µl 600 µl 120 µl |
0.50 % 1.00 % 2.50 % |
0.20 % 0.30 % 1.00 % |
| eLINE Electronic Pipette | 12 | 0.2-10 µl | 0.05 µl | 10 µl 5 µl 1 µl |
0.90 % 1.50 % 4.00 % |
0.50 % 0.80 % 3.00 % |
| eLINE Electronic Pipette | 12 | 5-120 µl | 0.50 µl | 120 µl 60 µl 12 µl |
0.50 % 0.70 % 2.00 % |
0.20 % 0.30 % 1.50 % |
| eLINE Electronic Pipette | 12 | 10-300 µl | 1.00 µl | 300 µl 150 µl 30 µl |
0.50 % 0.70 % 2.00 % |
0.20 % 0.30 % 1.00 % |
| eLINE Electronic Pipette | 12 | 50-1200 µl | 5.00 µl | 1200 µl 600 µl 120 µl |
0.50 % 1.00 % 2.50 % |
0.20 % 0.30 % 1.00 % |
* Note: Minimum volume in Pipetting mode is 500µl. The 100 µl is possible in Multiple Dispensing mode.
X: 0 = Without AC adaptor, 1 = Universal AC -adaptor(incl.Euro, US/Jpn, UK and China plugs)
* Note: Minimum volume in Pipetting mode is 500µl. The 100 µl is possible in Multiple Dispensing mode.
eLINE 0.1-5 µl with Super Pipetting
Type: brochure
eLINE electronic Pipette
Type: brochure
eLINE manual in English
Type: manual
eLINE manual in Finnish
Type: manual
eLINE manual in French
Type: manual
eLINE manual in German
Type: manual
eLINE manual in Italian
Type: manual
eLINE manual in Spanish
Type: manual
eLINE PCR Starter kit brochure
Type: brochure
The benefits of electronic pipetting - How to choose the right pipette?
Type: article